WATER QUALITY--Interim procedures for measuring pH in low-conductivity waters. June 23, 1980 QUALITY OF WATER BRANCH TECHNICAL MEMORANDUM NO. 80.19 Subject: WATER QUALITY--Interim procedures for measuring pH in low-conductivity waters. Districts are being asked with increasing frequency to measure pH in samples of precipitation and other waters of low conductivity. At least two effects are being observed in such samples that do not occur in most others; l) a delay of several minutes in reaching a stable reading in the sample and 2) large differences (up to 0.6 units) in the apparent pH between a stirred and a quiescent sample. Both of these effects have been discussed in a recent paper by Galloway and others (Limnology and Oceanography, 1979, vol 24, no 6, pp 1161-65). The first effect is probably caused by incomplete washing of the buffer from the glass electrode or by storing electrodes in buffer solutions. The second effect may be due to distortions of the potential field in the sample as a result of stirring, an effect which has been referred to (perhaps incorrectly) as a streaming potential. Both effects are thought to become insignificant at conductivities above 50 umhos, although this limit is not well established. The Quality of Water Branch is continuing to investigate the effects described above and to develop standard procedures for pH measurement under these conditions. Until such procedures are developed and tested, we advise all districts to: l) where possible use probes designed for low conductivity waters; and 2) make pH measurements according to the following interim procedure: 1. Calibrate using pH 7 and 4 or 9 buffers to bracket the expected pH. 2. Rinse the probes copiously with distilled water; do not wipe the probes. 3. Place probes in distilled water (or sample if plenty is available) with stirring for 10 to 20 minutes, replacing the water at least 3 times. During this step, periodically stop the stir and note the pH in quiescent water. When readings in quiescent water drift less than 0.1 units/minute, the system is ready to use. 4. Remove the probes, shake off excess water, gently pour the sample into a clean beaker, insert probes and gently swirl the sample to achieve good contact of sample to probes. 5. Make readings in a quiescent sample. Take the first reading after drift is less than 0.1 units/minute. 6. Remove the probes and rinse with distilled water. Measurements on several samples may be made from one calibration, skipping steps 1 through 3. If the meter must be moved, place the probes in distilled water and leave the meter in a measurement mode. The stability of calibration will vary with instrument and handling. The decision to recalibrate must be left to the judgement of the person making the measurements. Probes should be stored in distilled water, not buffer. Measurements with in situ probes, which will often be in flowing streams, should be checked against measurements in quiescent samples following the above procedure. When differences exceed 0.2 units, accept the measurements in quiescent samples. R.J. Pickering Memoranda superseded: None Keywords: Water quality, analytical methods, instrumentation, pH. WRD Distribution: A, B, S, FO, PO