EQUIPMENT & SUPPLIES--Marsh/McBirney model 201 electromagnetic meters; need to check manufacturer's calibrations
In Reply Refer To: November 13, 1980
EGS-Mail Stop 415
SURFACE WATER BRANCH TECHNICAL MEMORANDUM NO. 81.02
Subject: EQUIPMENT & SUPPLIES--Marsh/McBirney model 201 electromagnetic
meters; need to check manufacturer's calibrations
A check on the rating calibration of a Marsh/McBirney 201 current meter
in the towing tank at GCHC suggests the need for individually rating these
meters. Whereas the meter registers zero when there is zero flow, the
manufacturer's linear rating equation has a finite intercept. The actual
meter readings will therefore depart from the linear equation at the low
velocities, sometimes resulting in sizeable errors. Use of the Marsh/
McBirney 201 meter for accurate results will require that each meter be
calibrated individually. Meters to be rated should be sent according to
the standard procedures to the Repair and Calibration Section, Hydrologic
Instrumentation Facility, GCHC, NSTL Station, MS 39529.
Stanley P. Sauer
Acting Chief, Surface Water Branch
Distribution: A, B, S, FO, PO