Report: Evaluation of T-Hydronics TH-LD Submersible Pressure Transducers

To: "E  - All WRD Employees"<distribution@usgs.gov>
Subject: Report: Evaluation of T-Hydronics TH-LD Submersible Pressure Transducers
Mime-Version: 1.0
Content-Type: text/plain; charset="us-ascii"
Date: Fri, 10 May 1996 15:58:36 -0500
From: "Dorothy E Greenwood, Secretary (Steno), Bay St. Louis, MS" <degreenw>


Evaluation of T-Hydronics TH-LD Submersible Pressure Transducers



[NOTE: This evaluation is intended for internal USGS reference purposes only 
and does not constitute acceptance or rejection of the brand name product on 
the basis of the results obtained from the actual unit tested.]

The HIF recently completed an evaluation of two T-Hydronics Model TH-LD 
submersible pressure transducers. The purpose of the evaluation was to 
determine whether these devices meet manufacturer specifications and WRD 
requirements.  The results are summarized in this report.

General Description

The T-Hydronics TH-LD pressure transducer is designed to determine the water 
level in a well or borehole.  It consists of a stainless steel housing (1.5 
inches in diameter by 2.75 inches long) that contains the transducer element 
and electronic circuitry and a PVC-jacketed vented cable.  Operating range 
of the units tested was 0 to 10 pounds per square inch (psi) (0 to 23.1 feet). 
[NOTE:  Other ranges from 0 to 5 psi through 0 to 500 psi are available.] 
These units also have temperature compensation for -18 to 65 degrees Celsius 
and an analog voltage output of 0 to 3 millivolts per volt (mv/V) of the 
operating voltage.  The manufacturer's stated combined error for nonlinearity, 
hysteresis, and repeatability for the pressure transducer is 0.5 percent of 
the full scale output. For a 0 to 23 foot (10 psi) pressure transducer, this 
is +/- 0.11 foot. Over pressure ratings are 1.5 times the full scale.  Nominal 
operating voltage is 10 volts ac or dc.

Test Procedures

Standpipe test procedures

The sensors were tested in a 100-foot standpipe facility located at the HIF. A 
Campbell Scientific, Inc. CR10 data logger connected to a Paroscientific PS-
2 was used as the controller for the cycling of the water level in the 
standpipe.  Based on the sensor being tested, the CR10 was programmed to raise 
and lower the water level in the standpipe in four- or five-foot increments 
over a range near that of the full-scale range of the sensor.  Three 
temperature sensors also connected to the CR10, monitor the temperature of the 
water in the standpipe during the test.

The T-Hydronics TH-LD sensors were suspended by their cables and placed 
approximately 22 feet from the top of the standpipe.  Since the T-Hydronics 
TH-LD sensors produce an analog voltage output, they were connected to a 
second CR10 with analog voltage inputs (the primary CR10 analog inputs were 
used by the standpipe temperature sensors). The secondary CR10 was 
programmed to excite the sensors with 2.5 volts ac (Campbell's ac bridge 
measurement) and collect and store the average of ten analog voltage 
measurements every 30 seconds.  This CR10 was also programmed to communicate 
with the primary CR10 as an SDI-12 sensor.  After completion of setup, the 
CR10 was set to begin automatic operation.  Data was collected over 
telephone modems during the test periods.  At the end of each sensor test 
period, the automatic operation was shut off and the water level again set 
to the midpoint of the range of water levels and the sensors checked for any 
offset in readings.  The test ran for six cycles from October 11 to 14, 
1995.  Ambient water temperature of the standpipe during the tests was between 
22.5 and 28.4 degrees Celsius.

Environmental chamber test procedures

Two T-Hydronics submersible pressure transducers model TH-LD were tested in 
the environmental chamber. A National Institute of Standards and Technology 
(NIST) traceable pressure standard was used to measure the standard applied 
pressure that simulated the water level. For one transducer, a standard 
pressure equal to 0 to 23 feet (at 20 different water levels) was applied at 
four constant temperatures.  For the other transducer the same pressures 
were applied at five constant temperatures.

The pressure transducer measurements were recorded using a Campbell CR10 
data logger and a SM 192 data logger.  A constant 12 volts was supplied during 
the environmental testing. The pressure transducer analog 3 mv/V output was 
translated to feet of water using linear regression to define the equation for 
each constant temperature.

Test Results

Standpipe test results

The summary of the test results of the transducers is displayed in table 1. 
Figure 1 showing graphs of the test results is available upon request.

Environmental chamber test results

At constant temperature--Figures 2 and 3 (available upon request) represent 
the measurement differences from the standard at constant temperatures of 
25, 0, 50, and -5 degrees Celsius for both of the transducers.  Table 2 
summarizes the results of the differences from the standard values for the two 
units at constant temperatures under pressure cycling.  The differences 
between standard values and measured values are grouped in two ranges: 0 to 10 
feet and 10 to 23 feet.   For S/N 34952, in the 0 to 10 foot range, the 
largest span of difference from a minimum difference to a maximum difference 
was -0.08 to 0.06 foot.  For S/N 35415, in the 0 to 10 foot range, the largest 
span from a minimum difference to a maximum difference was -0.11 to 0.13 foot.  
For S/N 34952, in the 10 to 23 foot range, the span from a minimum 
difference to a maximum difference was -0.10 to 0.08 foot.  For S/N 35415, 
in the 10 to 23 foot range, the span from a minimum difference to a maximum 
difference was -0.18 to 0.13 foot.

Zero Reading--For one unit the difference from the standard measurement, 
when referenced to atmospheric pressure, ranged from -0.06 to 0.06 feet of 
water. For the other unit the difference ranged from -0.10 to 0.01 feet of 
water.

Measurement hysteresis--For S/N 34952, the largest pressure measurement 
hysteresis was 0.03 foot. The largest pressure measurement hysteresis for S/
N 35415 was 0.07 foot.

Comments

When a pressure transducer is used for measuring water level, the following 
factors need to be considered: A regulated constant voltage supply is required 
to maintain the accuracy of reading for an analog voltage output transducer. 
The data recorder and pressure transducer need to be calibrated as a system 
and always used as a unit. This ensures that the measurement error of the 
transducer and the error of the data recorder analog to digital conversion are 
combined and known. If necessary, the proper linear regression equation can be 
programmed into the data recorder and used to help correct the measurement 
error of the system.

Conclusions

Standpipe conclusions

The differences between the TH-LD's and the PS-2 at the temperatures tested 
were in the range of -0.31 to +0.28 feet, mean values were in the range of -
0.10 to +0.04 foot.  There was no discernable hysteresis based upon rising 
or falling stage.

Environmental chamber conclusions

Accuracy--Test results indicate that measurement differences are as much as 
0.18 foot.  For both transducers in the 0 to 10 foot range, test results 
indicate that measurement differences are greater than 0.01 foot.  For the 10- 
to 50-foot range, measurement differences were greater than 0.1 percent of 
reading for both transducers. (See table 2.)

Zero Setting--When the pressure transducer is vented to atmospheric 
pressure, the measurement can not be set to a zero value.

Zero Reading--When repeatedly vented to atmospheric pressure, the 
transducers do not measure the same value. The values varied as much as 0.12 
foot.

Measurement hysteresis--During pressure cycling, the pressure transducers 
did not repeat the same measurement for both applications of the same standard 
value.

SDI-12 Communication--These pressure transducers do not have SDI-12 
communication. These units have analog millivolt output.



 Table 1. Standpipe Tests--The range of differences and mean of the 
differences between two model T-Hydronics TH-LD submersible pressure 
transducers and a PS-2, in feet of water.



PS-2 Stage,                S/N 34952                     S/N 35415 
feet            Mean         Range            Mean         Range

0.00            -0.04      -0.27 ~  0.18      -0.01      -0.20 ~ 0.19  
4.00            -0.05      -0.21 ~  0.18       0.00      -0.07 ~ 0.22  
8.00            -0.06      -0.25 ~  0.21       0.01      -0.06 ~ 0.28 
12.00           -0.02      -0.25 ~  0.18       0.02      -0.09 ~ 0.24 
16.00           -0.03      -0.26 ~  0.17       0.02      -0.11 ~ 0.23 
20.00           -0.07      -0.28 ~  0.12       0.03      -0.19 ~ 0.08 
16.00           -0.06      -0.25 ~  0.06       0.04      -0.08 ~ 0.13 
12.00           -0.08      -0.27 ~  0.03       0.03      -0.07 ~ 0.13  
8.00            -0.09      -0.28 ~ -0.02       0.02      -0.09 ~ 0.09  
4.00            -0.10      -0.31 ~  0.02       0.01      -0.16 ~ 0.06  
0.00            -0.06      -0.27 ~  0.12       0.01      -0.20 ~ 0.09




 Table 2. Environmental chamber test-- Minimum, maximum, and mean 
difference, in feet of water, between standard values and values measured by 
two T-Hydronics model TH-LD submersible pressure transducers in two water 
level ranges and at the listed temperatures (S/N, serial number; Min, minimum; 
Max, maximum; ft, foot) 
_____________________________________________________________________________
S/N,                        Temperature, in degrees Celsius     

Water level                 -20       -5       25       50        0


Sensor 34952   Min.        -0.06    -0.05    -0.05    -0.08       --         
0-10 ft        Max.         0.06     0.04     0.07     0.06       --
               Mean        -0.00    -0.00    -0.00    -0.00       --


Sensor 34952   Min.        -0.09    -0.07    -0.10    -0.04       -- 
   10-23 ft    Max.         0.08     0.05     0.08     0.04       --
-              Mean         0.00     0.00     0.00     0.00       --


Sensor 35415   Min.        -0.11    -0.10    -0.06    -0.04      -0.10
   0-10 ft     Max.         0.13     0.12     0.08     0.04       0.11
               Mean        -0.01    -0.01    -0.01    -0.00      -0.01


Sensor 35415   Min.        -0.18    -0.17    -0.12    -0.06      -0.16
  10-23 ft     Max.         0.13     0.12     0.09     0.06       0.11
               Mean         0.01     0.01     0.01     0.00       0.01


For more information on the standpipe testing call Carl Scott (CTSCOTT) at 
(601) 688-1538, or for environmental chamber testing information, contact 
Trudy Olive (TEOLIVE) (601) 688-1558. 

