EQUIPMENT AND SUPPLIES--Modification of the field calibration chamber for the YSI dissolved oxygen meter
March 7, 1979
QUALITY OF WATER BRANCH TECHICAL MEMORANDUM NO. 79.08 (B)
Subject: EQUIPMENT AND SUPPLIES--Modification of the field
calibration chamber for the YSI dissolved oxygen meter
This is to call your attention to a potentially significant source
of error in measurement of dissolved oxygen with the YSI model
5739 dissolved oxygen probe used in conjunction with the YSI model
5075 field calibration chamber. The problem occurs because the
pressure compensation diaphragm located on the side of the probe
body in this model will not fit inside the field calibration
chamber. As a result, when the chamber is submerged in water for
calibration, the pressure transmitted to the inside of the
membrane is hydrostatic pressure and the pressure on the outside
of the membrane is atmospheric pressure. (Fig. 1). This pressure
imbalance generally results in a high reading on the meter and
leads to incorrect meter calibration. Figure 2 shows the potential
error that can be encountered by using the probe with an
unmodified calibration chamber.
To remove this potential error, a modification kit for the field
calibration chamber has been designed and is available at no cost
from the Operations Section, Reston, VA. The kit extends the air
chamber 2 inches so that the pressure diaphragm is enclosed within
the chamber during calibration (Fig. 3). The pressure is then
equalized and no calibration error will occur. No other
modification is needed except the rubber stopper is moved up the
probe housing to the smooth area above the manufacturer's logo.
Modified calibration chambers can still be used in conjunction
with the older style dissolved oxygen probes that do not have the
pressure compensation diaphragm feature. However, it is
recommended that all calibration chambers be modified even though
no new probes are in use at the present time.
The modification kit consists of three pieces: a 2 inch extension
for the chamber, an adapter sleeve for connecting the extension to
the original section, and a retainer ring that mounts above the
first ring. The components should be silver soldered to the
original chamber as diagrammed in Figure 3. The adapter sleeve
soldering must be watertight. To order modification kits, write to
Wayne Hammond, Operations Section, MS 405, Reston, VA, or
telephone, FTS 928-6861. There is no charge for the kits but the
districts must do their own modification work.
The YSI corporation has been notified of this calibration problem
and hopefully new calibration chambers will be of sufficient
length to enclose the pressure diaphragm within the chamber during
calibration.
R. J. Pickering
Chief, Quality of Water Branch
WRD Distribution: A, B, S, FO, PO