6.6.7 REPORTING

TWRI Book 9
(REVISED 9/2001)

Report alkalinity and ANC to three significant figures. Only the value from filtered samples is to be published as alkalinity. Titration values from unfiltered samples are to be entered and published as ANC.

Alkalinity and ANC should be reported in milliequivalents per liter (or microequivalents per liter), if possible. If this option is not available in the data base, calculate alkalinity and ANC in milligrams per liter, assigning all neutralizing capacity to the carbonate system.

If calculating ANC, alkalinity, bicarbonate, or carbonate in milligram units, then report:

Arrow Less than 1,000 mg/L, to whole numbers;
Arrow 1,000 mg/L and above, to three significant figures.
Arrow Carbonate alkalinity usually is reported in the data base in milligrams per liter as calcium carbonate.

Conversion factors listed below are taken from Hem (1985).

Multiply By To obtain

Alkalinity (mg/L as HCO3)

0.8202

Alkalinity (mg/L as CaCO3)

Alkalinity (mg/L as HCO3) 0.4917 Alkalinity (mg/L as CO32–)
Alkalinity (mg/L as CaCO3) 0.01998 Alkalinity (meq/L)
Bicarbonate (mg/L as HCO3) 0.01639 Bicarbonate (meq/L)
Bicarbonate (mg/L as HCO3) 16.389 Bicarbonate (µeq/L)
Carbonate (mg/L as CO32–) 0.03333 Carbonate (meq/L)
Hydroxide (mg/L as OH) 0.05880 Hydroxide (meq/L)

Report the average value for duplicate samples or the median when more than two replicate samples are used for quality control and the value falls within the appropriate quality-assurance criterion for variability.

Use the correct parameter code to indicate (1) the method of titration or calculation and (2) a filtered or unfiltered sample.


Field Manual Contents

Table of Contents Chapter 6
Table of Contents
Section 6.6

Back to
Section
6.6.6


 

Accessibility FOIA Privacy Policies and Notices

Take Pride in America logo USA.gov logo U.S. Department of the Interior | U.S. Geological Survey
URL: http://water.usgs.gov/owq/FieldManual/Chapter6/section6.6/html/section6.6.7.htm
Page Contact Information: Water Webserver Team
Page Last Modified: Monday, 18-Jun-2007 15:50:37 EDT