
Water Resources--Office of Water Quality
This document is also available in pdf format:
Chapter 6.7.2.pdf
6.7.2
CALIBRATION
Follow the manufacturer's instructions for instrument calibration and
record calibration readings and adjustments in the instrument log book.
|
Calibration of turbidity instruments against a Formazin or other
approved primary standard usually is done in the laboratory, with
instrument checks performed in the field. Use standards that bracket the
range of turbidity anticipated in environmental samples, if
possible. |
| For instruments that are factory calibrated in
standard turbidity units, the calibration procedure checks the accuracy of
calibration scales provided by the manufacturer. |
| Periodically check the accuracy and precision of
your instrument against that of another instrument. |
| Consult the manufacturer if the precision of your
instrument deviates 5 percent or more from the manufacturer's
specifications. |
The USEPA specifies that the turbidimeter must be calibrated with a primary
standard (a Formazin or a styrene divinylbenzine polymer standard such as
Amco AEPA-1 PolymerTM) (U.S. Environmental
Protection Agency, 1994). A solid scattering standard provided by the
manufacturer for setting overall instrument sensitivity for all ranges
should not be relied on unless the turbidimeter is demonstrated to be free
of drift on all ranges (U.S. Environmental Protection Agency, 1979).
Temperature changes affect Formazin turbidity standards and the performance
of the turbidity instrument.
| Turbidity
instruments are not currently available with an automatic
temperature-compensating function. |
| Standards and instruments should be at the same and
constant temperature during calibration to achieve stable and accurate
results. |
| To avoid the effects of thermal fluctuations on the
calibration, perform the Formazin calibration and calibration of the
secondary standard (for example, GelexTM)
against the primary standard in the office laboratory at room temperature
instead of at the field site. At the field site, check instrument
calibration using a secondary standard. |
Preparation of the stock turbidity suspension and standard
dilutions
Prepare the stock turbidity suspension monthly and standard dilutions on
the day of instrument calibration. To prepare and dilute a 400 NTU
Formazin stock suspension4:
- Dissolve 1.000 g hydrazine sulfate [(NH2)2
H2SO4] in
filtered water and dilute to 100 mL in a volumetric flask.
- Dissolve 10.00 g hexamethylenetetramine [(CH2)6N4] in filtered water and dilute to 100 mL in a
volumetric flask.
- Mix 5.0 mL of hydrazine sulfate and 5.0 mL of hexamethylenetetramine
solutions in a 100-mL volumetric flask and let stand 24 hours at 25 ±
3°C; dilute to the mark and mix. To prepare 500 mL of 400 NTU
standard, mix 25 mL of the reagent solutions in a 500-mL flask, dilute to
the mark, and mix.
- For a 40 NTU standard, dilute 10.00 mL of the 400 NTU stock suspension
to 100 mL with turbidity-free water (sample or deionized water passed
through a filter media of
0.2
µm).
- Dilute stock suspension on the day the standard is needed, use it
immediately after preparation, and discard unused standard.
- Inconsistent techniques used to dilute standards can add as much as 5
percent measurement error.
6.7.2.A
TURBIDIMETER CALIBRATION
The calibration instructions and procedures that follow are general and
should be modified to apply to the instrument being used--check
manufacturer's instructions:
- Prepare Formazin suspensions as described above.
- Calibrate each instrument range using at least three standard
concentrations. Use standards that bracket the range of turbidity
anticipated in the sample solution.
- Prepare dilute standards fresh from the stock at the time of use--after
dilution, the stock suspension is stable only for 4 to 6 hours.
- For turbidity greater than 40 NTU, use undiluted stock solution.
- Do not use standards with flocculated
suspensions.
- Switch the turbidimeter on and allow it to warm up. Put on disposable
gloves.
- Check instrument focus: insert template in the cell holder. The lamp
image should just fill the inside circle. Adjustment is required if the
image is off center, too large, or too small.
- Field rinse a clean, dry, scratch-free, index-marked cell with the
highest concentration of the standard for the instrument range setting or
range of interest.
- Hold the sample cell by the rim (top lip), not beneath the
lip.
- Pour standard into the sample cell to the fill mark.
- Wipe the exterior of the cell using a soft, lint-free cloth or tissue
to remove moisture (condensation) on cell walls.
- Apply a thin layer of silicon oil (table 6.7-1) onto the exterior of the cell
to reduce condensation on the cell and mask slight scratches and nicks.
Apply silicon oil uniformly onto the blank cell if it will be used on the
cell filled with standard (check manufacturer's recommendations).
- Select the desired NTU range.
- Set the calibration adjustment to equal the high value of standard for
the range of interest.
- Before inserting the standard, ensure that no air bubbles are
present.
- Orient the standard cell in the cell holder--the calibration cell and
sample cell must have identical orientation when in the instrument
measurement chamber.
- In the instrument log book, record and graph the instrument value for
each standard (instrument reading versus standard value--see fig. 6.7-1).
- Adjust standardization control until the value on the meter equals the
NTU value of the standard used.
- Remove the sample cell and discard the first turbidity standard.
- Rinse and fill a clean cell with the second turbidity standard and
orient the cell in the instrument.
- Take a reading without adjusting the calibration.
- Plot this instrument NTU reading against the NTU value of the turbidity
standard (fig. 6.7-1).
- Repeat step (9) for at least one more turbidity standard with NTU value
to cover the turbidity range of interest. The greater the number of
turbidity standard values used, the greater the reliability of the
calibration.
- Prepare a calibration curve for each range of values to be used if a
precalibrated scale is not supplied by the manufacturer. (The accuracy of
calibration scales provided with the instrument must be verified by using a
precalibrated instrument and appropriate standards.)
- The plot of instrument reading versus turbidity standard value is a
range calibration curve.
- Verify that any instrument reading (dial setting) within the range
calibrated is correct and agrees with correlative points on the calibration
curve.
- Calculate the NTU of a diluted sample:
NTU = A x (B+C) / C
where
A = NTU found in diluted sample,
B = volume of dilution water, in milliliters, and
C = sample volume taken for dilution, in milliliters.

6.7.2.B
SUBMERSIBLE TURBIDITY SENSOR
CALIBRATION
Most multiparameter instruments with turbidity probe capability are
microprocessor-based, with the calibration parameters stored in instrument
memory. Turbidity values of the standards are user-selectable in some
instruments, but some instruments have internally established standard
values that cannot be changed. Low-level check standards in the 1-5 NTU
range will allow the user to assess the actual performance of the
instrument near the detection limit; instrument reliability
generally decreases at NTU less than 5--consult manufacturer's
specification for the expected accuracy of the measurement.
Monitor digital output carefully to assure that turbidity readings are
stable before confirming the calibration. Note that if the instrument uses
signal averaging to smooth instrument output, output response to changes in
turbidity readings can be slowed.
Calibrate the instrument before leaving for the field
site. While in the field, check instrument performance
periodically using turbidity standard and turbidity-free water. The
optical surface of the probe must be clean before beginning the calibration
procedure. Modify the general instructions that follow as necessary so
that they are compatible with the manufacturer's instructions:
- Prepare a sufficient volume of the Formazin standard, as described
previously. Volume of standard required could be 500 mL for some
instruments, particularly if the entire sonde bundle instead of just the
turbidity probe will be immersed.
- Select Procedure (A) or (B). The same procedure, once
tested and selected, also should be used in future studies.
Procedure A. Immersion of entire sonde (bundle of
field-measurement sensors, including the turbidity sensor)--requires larger
volumes of standard; standard is vulnerable to contamination and dilution.
The sonde sensor guard may need to be removed.
Procedure B. Immersion of turbidity probe only--depending
on sonde configuration, isolation of the turbidity probe and achieving a
bubble-free optical surface could be difficult. This technique minimizes
the volume of standard required for calibration.
- Using a zero NTU standard (turbidity-free water):
- Rinse sonde/probe with deionized water, followed by a portion of
turbidity-free water.
- Immerse entire surface of sonde/probe in turbidity-free water.
- Agitate the sonde/probe repeatedly to remove bubbles from the optical
surface (activate mechanical wiper, if present).
- Monitor turbidity readings for 1 to 2 minutes or longer to ensure that
readings are stable (consult manufacturer's recommendations and
signal-averaging information).
- Confirm the zero NTU calibration point using manufacturer's
instructions.
- Remove sonde/probe and dry thoroughly to minimize dilution or
contamination of the next standard.
- Discard first standard (turbidity-free water).
- Using the second standard (Formazin suspension):
- Rinse sonde/probe surfaces with a portion of standard. Discard
rinsate.
- Immerse entire surface of sonde/probe in a container filled with
standard.
- Agitate the sonde/probe repeatedly to remove bubbles from the optical
surface (activate mechanical wiper, if present).
- Monitor turbidity readings for 1 to 2 minutes or longer to ensure that
readings are stable (consult manufacturer's recommendations and
signal-averaging information).
- Confirm the NTU calibration point for the standard used, according to
manufacturer's instructions.
- Remove sonde/probe and rinse surfaces thoroughly with deionized water
followed by turbidity-free water. Dry sonde/probe thoroughly.
- Discard used standard.
- Repeat steps 4(a-g) using a different Formazin suspension standard if
increased accuracy is desired and instrument software permits use of a
third calibration point.
- By diluting the existing standards, prepare a standard with turbidity,
either approximately midway between the calibration points and (or) close
to the estimated turbidity of the water to be measured.
- Measure the turbidity of this suspension, making certain that it is
within the accuracy specification of the instrument with regard to the true
value.
- Repeat the calibration procedure if the measurement is not within the
specification.
Once the instrument is calibrated, the accuracy of the recorded
measurements can be increased by preparing a calibration graph using
dilutions of the Formazin standards, as described previously for
calibration of turbidimeters (6.7.2.A).
TECHNICAL NOTE: Multiparameter
instruments with turbidity-probe capability use a light-emitting diode in
the range of near-infrared wavelength as the radiation source and usually
are microprocessor-based. The USEPA has not approved instruments using
this method as of this writing, and the accuracy attainable with
probe-based instrumentation is substantially less than that of
USEPA-approved instruments. ISO turbidity-measurement criteria were
developed to improve measurement consistency of instruments using the near
infrared technology, and some of the field instruments available meet ISO
7027 recommendations (table
6.7-2).
6.7.2.C
SPECTROPHOTOMETER CALIBRATION
Spectrophotometric turbidity measurements are useful to indicate relative
values or to monitor changes in turbidity with time. Spectrophotometers
are inaccurate for absolute turbidity values, and the instrument
sensitivity is unrated.
Spectrophotometers commonly have a stored program for turbidity that has
been factory-calibrated. This can be verified but not adjusted. Check the
instrument output against that of a different instrument every few weeks
while the instrument is in use. Check the relative accuracy of the
turbidity measurement before leaving for the field by inserting Formazin
standards covering the FTU range needed.
- Use freshly prepared standards.
- Be accurate in your dilution of the stock suspension.
- Prepare standards daily and discard any unused portion after each
use.
- Wear disposable powderless (vinyl or latex)
gloves--fingerprints or smudges on cuvettes cause false turbidity
readings; oils from skin can etch the cuvette glass.
- Hold the sample cell (cuvette) at the rim (on the top lip), not beneath
the lip. Pour standard into sample cell to the fill line.
- Wipe the exterior of the sample cell with a clean, soft, lint-free
cloth or tissue after filling to remove moisture and condensation from cell
walls.
- Check periodically for condensation on the sample cell and wipe it
dry.
- After wiping condensation from cell walls, apply a light coating (two
drops) of silicon oil (optical grade) using a lint-free cloth--check
recommendations from the instrument manufacturer.
- Eliminate gas bubbles from standards.
- Check that the calibration cell and sample cell have the same
orientation when placed into the instrument measurement chamber.
4Refer to American Public Health Association and others
(1992) for detailed instructions.
Section 6.7.3
Return to Section 6.7.1
Return to Contents for 6.7--Turbidity
Return to Chapter A6 Contents Page
Return to Field Manual Complete Contents
Return to Water Quality Information Pages Maintainer: Office of Water Quality
Webversion by: Genevieve Comfort
Last Modified: 22JUNE98 ghc